ISO 9001:2015 and ISO/IEC 17025:2017 accredited

ISO 9001:2015 and ISO/IEC 17025:2017 accredited

  • Nationwide Service

  • 24/7 Emergency Service

  • Calibration & Service Experts

  • 45+ Years Canadian-Owned

A&D

Moisture Analyzers - Models MS-70 / MX-50

Service & Calibration Support

Interweigh Systems provides calibration, service, and technical support for all equipment we sell.

Our technicians service equipment nationwide and can support commissioning, preventative maintenance, troubleshooting, and re-certification as required.

Certified test weights are also available to support quality control programs and internal verification procedures.

Ordering & Configurations

Most equipment is configured based on application requirements. Pricing and lead times may vary depending on configuration, accessories, and certification needs.

To receive pricing or discuss availability, contact our team with details about your application and environment.

Contact Now for Immediate Pricing

Description

Moisture Analyzers - Models MS-70 / MX-50

Moisture Analyzers from A & D Weighing offer superior moisture detection and sensors with the highest measuring accuracy and functionality available. Designed specifically for the detection of moisture in the laboratory and many other related applications, the MS-70 and MX-50 models have a weighing capacity of 51 g to 71 g and are capable of measuring moisture content at 1% to 0.001% resolution suitable for low moisture content samples.

For more information download the Moisture Analyzers brochure.

Notable Features

  • Fast and uniform heating with halogen lamp and uniquely designed SRA (Secondary Radiation Assist) filter.
  • WinCT-Moisture Software (standard with MS & MX, option with MF) displays a real-time graph of moisture content rate change.
  • Sodium Tartrate Dihydrate is provided as a stable moisture content for accuracy checks of the analyzer.
  • Five Measurement Programs: Standard. Automatic. Timer, Manual, Quick.
  • Clear and easy-to-see VFD display
  • Glass window for heating check and easy cleaning.
  • Self-check function.
  • Quick Reference Card.
  • Standard bi-directional RS-232C
  • Conformity to GLP, GMP, GCP and ISO with time/date, ID, calibration data and check record outputs.
Model Max Sample Weight Capacity Weight Resolution Moisture Content Display Moisture Content Accuracy - over 1 gram Moisture Content Accuracy - over 5 grams
MS-70 71 g 0.0001 g 0.001/0.01%/.1% 0.05% 0.01%
MX-50 51 g 0.001 g 0.01/0.1% 0.05/0.1/1% 0.10% 0.02%
Measurement Method 400 Watt straight halogen lamp heating system with SRA filter and SHS weighing technology
Heating Technology Halogen lamp (straight type, 400 Watt max, 5000 hours
Drying Temperature 50 ~200°C (1° increment)
Measurement Program Memories 20 sets
Data Memory 100
Measurement Programs Standard / Automatic / Timer / Manual
Measurement Modes Wet base / Dry base / Solid content / Ratio
Display Type VFD
Interface Bi-directional RS232
Communication Software Win CT Moisture Standard
Operating Temperature 5 ~ 40°C / 40 ~104°F
Self Check Function Standard
Sample Pan Size .85 mm / 2.16"
Physical Dimensions/Weight 215mm(W) x 320mm(D) x 173mm(H)
8.46"(W) x 12.6"(D) x 6.8" (H)
Weight 13 1bs